Features:

  • Steam sterilization and CIP capability
  • 3m cable
  • 316 Stainless Steel
  • Sanitary Flange designed for Ladish or Tri-Clover fittings
  • O-ring seals used on all versions for high on-stream reliability
  • Wide range of cell constants 0.01, 0.02, 0.05, 0.1, 1.0, 2.0, 3.0, 5.0, or 10.0
The Turtle Tough Conductivity sensors are welded to custom size blind flanges and can then be inserted into standard Tee fittings in sanitary systems using fittings by Ladish or Alfa-Laval Tri-Clover. They are ideal for chemical CIP inline installation to directly replace existing toroidal installation schemes. Only FDA compliant food grade materials are used. Flange sizes include 1/2″, 1″, 1.5″, 2″ and 2.5″. Rated for use up to 150 psig at 130 degree Celsius. Cell Constants from 0.01 to 10.0. Wetted materials of construction are 316 stainless steel and Teflon or PEEK, with double redundant EPR O-ring seals (Viton optional). The process side O-ring is the only one in contact with the stream, allowing the back O-ring to maintain reliable sealing, free from chemical attack. TC elements are potted in the body itself. Cable is 3 meters standard but extended lengths are available. Process connections are customer specified sanitary flanges. The insertion depth is tailored for each flange size to obtain proper positioning in the standard tee fittings of the given piping system. The sensor constructions provide great flexibility in cleaning options including steam sterilization, or cleaning by solutions of bleach, ozone, or caustic.